Designed
for e.g. Private Clinics, Smaller Hospitals, Emergency
Departments, First Aid Units, Intensive Care Units
etc. Its main application focus is in Abdominal, OB/GYN
and Vascular Imaging. Powerful Color and Spectral
Doppler technology, high quality and high resolution
of tissue visualization provide maximum productivity
and diagnostic accuracy. |
|
|
|
|
| |
TECHNICAL
SPECIFICATION:
|
|
|
|
|
| |
1. DISPLAY MODES:
- B, B+B, 4B,B+Zoom, Zoom, M, B+M
- Option: PW (Pulse Wave Doppler)
- Option: Color Doppler: CF (Color Flow, PF (Power
Flow ),
B+CF/PF+PW
|
|
|
|
|
| |
2. SCANNING MODES:
- electronic linear / convex / micro-convex |
|
|
|
|
| |
3. FEATURES:
- color
Doppler module (option)

- SAS - Spectral Adaptive
System
- THV - Tissue Harmonic Visualization
- ESR - Enhanced Speckles Reduction
- Dynamic aperture, apodization & mix-dynamic
focusing
- 5 zones of TGC control for each of the scan depth
- Gray scale 256 shades of gray, image memory 512 x
512 x 8
- Display: Hi-Res non - interlaced 15" LCD
monitor
- Scanning frame rate up to 25 fps (automatically
optimized)
- M-mode sweep speed: selectable from about 2 to 9
sec/screen
- Zoom 6 steps
- Scanning depth range: from 2 cm to 25 cm
- Image orientation: L/R, Up/Dn, Pos./Neg.
- Cineloop memory (256-frame)
- Digital filters - adaptive Correlation, Edge
Enhancement, Wall filter,
Reject
- Gray scale postprocessing
- Image annotation: alphanumeric, comments, patient
information,
date, time, body marks
- Large, versatile trackball
- Independent report pages: B, M, PW, OB/GYN
- Probe interface: 1x electronic (+ 1 for special
probes - option)
- Easy upgrade and customize to specific diagnostic
requirements |
|
|
|
|
| |
4. OPTIONAL EQUIPMENT:
- videoprinter
B/W


- videoprinter
color


- biopsy
attachments 
- Internal
non-volatile memory module up to 1000 images
capacity

- USB-Pen
Memory Module up to 2000 images capacity
(transfer images to PC using USB Pen Drive)

- Internal or car battery supply |
|
|
|
|
| |
5. MEASUREMENTS:
B-mode:
- Distance (4 pairs of cursors)
- Area, circumference (ellipse, contour)
- Volume (ellipsoid, contour, 3-axis method)
- Urology calculations and others according to order
Prenatal calculation programs:
- BPD, CRL, FL, GS, AC, HC (factory & user table)
- fetal weight FW [AC/BPD]
- ratios calculation FL/BPD, HC/AC, FL/AC
- GA mean (automatic averaging of fetal age)
- automatic calculation of the date of birth EDD
- user and manufacturer table
M-mode:
- multi-measurement possibility
- time, velocity, amplitude (t, v, d)
- Heart rate (HR)
- LV measurement calculation: EDV, ESV, CO, SV, EF
- fast measurements of: AOD, LA, RVED, DE_Amp, IVS
Doppler PW:
- time, speed, acceleration (t, v, a)
- PHT, MVA, max. PG, mean PG, HR
- vessel index: A/B, RI, PI
|
|
|
|
|
| |
6. MULTI-FREQUENCY TRANSDUCER TYPES:
Electronic:
-
convex: CA-255 /R60
(2,5~5,0 MHz) 
- convex: CA-305/ R20
(2,5~6,0 MHz) 
- convex: CA-307/ R13
(3,0~7,0 MHz)
- convex: CA-409/ R13
(4,0~9,0 MHz)
- linear: LA-575 /L64
(5,0~7,5 MHz) 
- linear: LA-510 /L40
(5,0~10,0 MHz) 
- linear: LA-912 /L30
(9,0~12 MHz) 
- linear endorectal:
LR-575/L64 (5~7,5 MHz)

Mechanical:
- convex: S-255
/R15(2,5~5MHz) 
- convex: S-510
/R15 (5~10MHz) 
- micro-convex:
SM-457 /R10 (4,5~7MHz)

- micro-convex:
SR-457 /R10 (4,5~7 MHz)

|
|
|
|
|
| |
7. POWER SUPPLY:
- Power supply: AC
100/240 V, 50/60 Hz
- Power consumption: 40 VA |
|
|
|
|
| |
8. DIMENSIONS AND WEIGHT:
- 455 x 445 x 93
mm; 8,5 kg*
- trolley: 465 x 512 x 775 ca. 5,5 kg
* - depends on configuration
|
|
|
|
|
|
ECHO-SON reserves the right to
change design and specification without notice. All
rights reserved. |
|
|
|
|
|
|
|
|
|
|